Your Location: Home > English > Products > Electric instrument and DC power supply > Bench-multimeter >

VA83

  • Product Name :
    6 1/2 Bench Multimeter
  • Product Model :
    VA83
  • Product Series :
    Electric instrument and DC power supply
  • Product Brief :
    Key Features 12 different measurement functions: DCV/ACV, DCI/ACI, Ω2W/Ω4W, frequency/period, diode test, continuity, dB/dBm, etc. High-brightness vacuum fluorescent display (VFD) True-rms AC voltage a...
Key Features
12 different measurement functions: DCV/ACV, DCI/ACI, Ω2W/Ω4W, frequency/period, diode test, continuity, dB/dBm, etc
Maximum reading of 120,000 digits
DCV measurement accuracy up to 0.0035%, resolution 1µV DCV
True-rms AC voltage and current measurement, with high frequency of 300kHz
Maximum measurement rate: 45times/second
Equal accuracy frequency measurement more than 1MHz, minimum resolution of 10μHz
Good reading stability, high accuracy and high measurement rate
High-brightness vacuum fluorescent display (VFD)
Zeroing (RFL) mode to eliminate residual reading
2W and 4W resistance measurement modes selectable
Built-in mathematics calculation functions: mX+b, %, dB, dBm
512 readings storage and MAX/MIN/AVER/STD statistics
HI/IN/LO comparator function
GPIB and RS-232 interfaces for convenient system communication
Calibration performed without opening the cover
10 sets of measurement states saved and recalled up
Specifications
Full-scale reading & reading rate (readings/second)
Speed Slow
 
Medium
 
Fast
 
Full-scale reading (digits) 1199,999
 
119,999
 
11,999
 
Reading
rate
(readings/
second)
DCV,DCI 3.6
 
13
 
45
 
ACV,ACI 2
 
3.9
 
22.7
 
Ω-2W 3.6
 
13
 
45
 
Ω-4W 2.7 10 32.9
DCV
Range
100mV
1V
10V
100V
1000V
Max. reading
119.9999
1.19999
11.9999
119.9999
1010.000
Resolution
0.1µV
1µV
10µV
1V
1mV
Accuracy (1 year)
0.0065+0.0045
0.0040+0.0009
0.0035+0.0005
0.0045+0.0006
0.0055+0.0015C
Input impedance
>10GΩ
>10GΩ
>10GΩ
10MΩ
10MΩ
DCL
Range
10mA
100mA
1A
3A
 
Max. reading
11.99999
119.9999
1.199999
3.03000
Resolution
10nA
0.1µA
1µA
10µA
Accuracy (1 year)
0.050+0.004
0.050+0.004
0.080+0.004
0.120+0.015
Load voltage/shunt resistance
<0.15V/10Ω
<1.5V / 10Ω
<0.3V / 0.1Ω
<1V / 0.1Ω
AC voltage (ACV true-rms measurement)
Range 10mA 1V 10V 100V 750V
Max. reading 11.9999 1.199999 11.99999 119.9999 757.50
Resolution 0.1µA 1µV 10µV 0.1mV 1mV
10~20 Hz 1.50+0.20
20~50 Hz 0.50+0.10
50~100 Hz 0.10+0.03
100~20 kHz 0.05+0.03 0.08+0.03
20~50 kHz 0.15+0.05 0.11+0.05 0.18+0.05 --------
50~100kHz 0.60+0.08 --------
100~300kHz 4.00+0.05 -------- --------
AC voltage (ACV true-rms measurement)
Range 10mA 1A 3A
Max. reading 11.99999 1.199999 3.03000
Resolution 10nA 1µA 10µA
Accuracy
(1year)
10~20 Hz 1.50+0.10
20~50 Hz 0.50+0.03 0.60+0.05
50~100Hz 0.10+0.3 0.12+0.03 0.15+0.03
100~2 kHz 0.05+0.03 0.10+0.04 0.12+0.04
2~5 kHz 0.10+0.03 0.50+0.03 0.60+0.03
5~10 kHz 0.20+0.03 2.00+0.10 2.50+0.10
Load voltage/shunt resistance <0.15V/10Ω <0.3V/0.1Ω <1V/0.1Ω
Resistance (Ω2W/Ω4W)
Range Max. reading Resolution Measurement Accuracy (1 year)
100.000Ω 119.9999 100µΩ 1 mA 0.010+0.004
1.00000 kΩ 1.199999 1mΩ 1 mA 0.010+0.001
10.0000 kΩ 11.99999 10mΩ 100μA 0.010+0.001
100.000 kΩ 119.9999 100m Ω 10μA 0.010+0.001
1.00000 MΩ 1.199999 10μA 0.010+0.001
10.0000 MΩ 11.99999 10Ω 7.0×Rx/(10M+Rx) 0.040+0.001
100.000 MΩ 119.9999 100Ω 7.0×Rx/(10M+Rx) 0.800+0.010
Frequency
Range Max. reading Range Accuracy (1 year) Sensitivity (sine wave)
5~10 Hz 9.999999 1μHz 0.050 200mV rms
10~100Hz 99.99999 10μHz 0.010 40mV rms
100~100 kHz 99.99999 10mHz 0.005 40mV rms
100k~1MHz 999.9999 0.1Hz 0.005 100mV rms